Duplex PNP200 Two-Stage Gas Manifold

Duplex PNP200 Two-Stage Gas Manifold

Manifolds and stabilizersGCE Healthcare

The Duplex PNP200 two-stage gas manifold from GCE Medical is intended for use in hospital pipeline systems as a medical gas source.

Talk to the experts

Duplex PNP200 Two-Stage Gas ManifoldIntroduction

The GCE Medical Duplex PNP200 gas manifold is designed for use in hospital pipeline systems as a reliable medical gas source. It efficiently reduces gas pressure in two stages, maintaining a consistent distribution pressure.

The Duplex PNP200 features an integrated third source entry point for LOX, a compressor, or a backup gas source. The innovative touchscreen alarm, included with the manifold, displays all three gas sources on the alarm screen for enhanced monitoring.

The automatic manifold delivers gas from the operating bank to the pressure regulator until the cylinders are depleted. Once the supply is exhausted, the system automatically switches to the reserve bank, allowing for seamless replenishment. This automatic switch is powered by a new, advanced pneumatic shuttle valve design.

Duplex PNP200 Two-Stage Gas ManifoldFeatures

All-in-One Control Panel

Manage all three gas sources seamlessly from a single panel to simplify operation and improve workflow efficiency.

Versatile Cylinder Compatibility

The manifold is compatible with 300 bar, 200 bar, and 150 bar cylinders to provide flexibility for various hospital setups.

Suitable for All Medical Gases

Designed to work with a wide range of medical gases for versatility.

Fully Automatic Changeover

Automatically switches to the reserve gas bank when the primary source is exhausted for uninterrupted gas supply without manual intervention.

Innovative Touchscreen Alarm

Equipped with a touchscreen alarm that provides real-time visual and audible notifications.

Real-Time Source Monitoring

The integrated alarm system continuously monitors all gas sources, displaying their status and offering visual cues for active sources.

Integrated Communication Protocols

Supports both MODBUS TCP/IP and RS485 RTU for easy integration with existing hospital systems and external monitoring devices.

Third Source Option

Includes an additional port for LOX, a compressor, or a third cylinder to provide flexibility for backup gas sources or specialised setups.

Leakage Monitoring for Safety

Monitors potential leakage on non-active sources, alerting users to any issues with the reserve side and enhancing overall system safety.

Maintenance Without Interruption

Perform service and maintenance operations while the manifold continues to function, ensuring no gas supply disruption during critical times.

Simple and Quick Installation

The manifold comes with a pre-installed back plate to simplify installation.

Optional Terminal Unit Test Point

Provides easy access for testing and servicing, ensuring the manifold operates efficiently and meets safety standards.

Duplex PNP200 Two-Stage Gas ManifoldGeneral properties

General properties

Gas type(s) O2, Air, Air800, N2O, CO2, (all medical gases)
Nominal flow200 m3/h
Cylinder inlet pressure300 bar (30 000 kPa) / 200 bar (20 000 kPa)
LOX or Compressor inlet pressure8 - 14 bar (800 - 1 400 kPa)
3rd source standby manifold pressure7 - 10 bar (700 - 1 000 kPa)
Outlet nominal pressure4,5 bar ( 450 kPa) (8 bar Air800)
Inlet connection2x W21,8×1/14“M, 1xM20x1F
Outlet connectionG3/4“M
Outlet pressure relieve valve6,8 bar (including manual activation) (12 bar Air800)
Intermediate pressure relieve valve17 bar (including manual activation)
Pressure relieve valve pipe dimensionø 15 mm
Testing pointNIST QC or QC in national standard
AlarmIntegrated with touch screen
Alarm power supply24VAC + added trafo 110V-240V
Alarm languagesEN, FR, SE, IT, messages in any language in latin letters can be set
Integrated communication protocolsMODBUS TCP/IP and RS485 RTU
Output relay2x NO/NC, 1A max
Regulatory statusMDD 93/42/EEC, EN ISO 7396-1, EN 60601-1-2, SIS HB 370

Configurations

Loading...
... loading table ...

Quality Standards

We comply with all relevant regulations

See our CERTIFICATIONS >

Duplex PNP200 Two-Stage Gas ManifoldDownloads

Find our Products

Source our products wherever you are

You can have confidence in our global network

Find a distributor
GCE main LogoPart of ESAB Corporation

GCE Group

Turmstrasse 30

6321 Steinhausen

Zug

Switzerland

Copyright © 2025 GCE. All rights reserved.
facebooktwitterlinkedinyoutubeinstagram